Table of Contents

Advertisement

www.allice.de
®
R&S
NGE100B

6.5.9 OPP Setting

User Manual 5601.1343.02 ─ 02
Example:
INST OUT1
VOLT:PROT ON
OVP of Ch 1 is activated.
VOLT:PROT? -> 1
Return the OVP state of the selected channel.
[SOURce:]VOLTage:PROTection:LEVel <Voltage>
This command defines or queries the OVP value of the previous selected channel.
Parameters:
<Voltage>
{<Voltage> | MIN | MAX}
<Voltage>
Set the OVP value in the range of 0 mV up to 32.2 V (adjustable
in 100 mV steps)
MIN
Set the OVP value to minimum.
MAX
Set the OVP value to maximum.
Default unit: V
Example:
INST OUT1
VOLT:PROT:LEV 5
Set the OVP value of Ch 1 to 5 V.
VOLT:PROT:LEV?
Return the selected channel OVP value.
[SOURce:]VOLTage:PROTection:CLEar
This command resets the OVP state of the selected channel. If the OVP has tripped
the OVP message on the display will be cleared for the selected channel.
Usage:
Event
[SOURce:]POWer:PROTection:LEVel <Power>
This command defines or queries the OPP value of the previous selected channel.
Allice Messtechnik GmbH
Remote Control
Remote Control - Commands
86

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nge102bNge103bNge100bNge-k102Nge-k103

Table of Contents