6.5.8 OVP Setting

User Manual 5601.1343.02 ─ 02
[SOURce:]VOLTage:PROTection:MODE <arg0>
This command defines or queries the OVP mode for the previous selected channel.
Parameters:
<arg0>
MEASured | PROTected
MEASured
The OVP switches off if the measured value exceeds the thresh-
old.
PROTected
If the adjusted threshold is exceeded the output of the instru-
ment will be not switched on; additionally the measured value is
monitored (please also refer to function MEASured).
Example:
INST OUT1
VOLT:PROT:MODE PROT
Sets OVP protected mode for Ch 1.
VOLT:PROT:MODE? -> PROT
Return the selected channel OVP mode.
[SOURce:]VOLTage:PROTection:TRIPped?
This command queries the OVP tripped status of the previous selected channel.
Return values:
<arg0>
ON | OFF
ON
OVP is tripped
OFF
OVP is not tripped
Example:
INST OUT1
VOLT:PROT:TRIP? -> OFF
OVP of Ch 1 has not tripped.
Usage:
Query only
[SOURce:]VOLTage:PROTection[:STATe] <arg0>
This command defines or queries the OVP state for the previous selected channel.
Parameters:
<arg0>
1 | 0
1
OVP is activated.
0
OVP is deactivated.
Allice Messtechnik GmbH
Remote Control
Remote Control - Commands