A beep is sounded if buzzer is enabled. See
Figure 5-13: Indicator of OVP/OPP when tripped
1. Select "Output > Protection" menu item.
The R&S NGE100B displays the "Protection" dialog. See
Figure 5-14: Protection dialog
2. Set the OVP mode for the desired channel.
The available options are as follows:
"Measured": The output is switched off if the measured value exceeds the set
limit. Switching on the channels triggered the OVP if the condition remains.
"Protected": The output is disabled if the setting exceeds the set limit.
"Disabled": OVP is set to maximum value supported by hardware. See the R&S
NGE100B brochure for the instrument specification.
3. Set the desired voltage for OVP.
You can set a value from 0 V to 32.5 V.
The R&S NGE100B turns off the respective channel when the measured voltage
exceeded the preset value.
4. Set the OPP mode for the desired channel.
The available options are as follows:
"Enabled": The channel is turned off if the measured power exceeds OPP
value.
"Disabled": Output power is limited to maximum value of 33.6W.
Allice Messtechnik GmbH
Instrument Functions
Advanced Functions
Figure
5-25.
Figure
5-14.