6.5.11 Voltage and Current Configuration Commands

User Manual 5601.1343.02 ─ 02
APPLy <arg0>
This command defines or queries the voltage and current value of the selected chan-
nel.
Parameters:
<arg0>
{<Voltage> | DEF | MIN | MAX} [, {<Current> | DEF | MIN | MAX}]
<Voltage>
Set voltage in the range of 0 V to 32.20 V.
DEF
Set voltage to default.
MIN
Set voltage to minimum .
MAX
Set voltage to maximum.
<Current>
Set current in the range of 3 mA to 3.025 A.
DEF
Set current to default.
MIN
Set currernt to minimum.
MAX
Set current to maximum.
Example:
INST OUT1
APPLY "6,2"
Set channel 1 to output 6 V and 2 A.
APPLY? -> 6, 2
Return voltage and current value of the selected channel.
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <Current>
This command defines or queries the current value of the selected channel.
Parameters:
<Current>
<Current>
Set the selected channel to a current range of 3 mA to 3.025 A
Default unit: A
Example:
CURR 2
Set the selected channel current to 2 A.
CURR? -> 2
Return current value of the selected channel .
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <Voltage>
This command defines or queries the voltage value of the selected channel.