www.allice.de
®
R&S
NGE100B
6.5.1 Common Commands
User Manual 5601.1343.02 ─ 02
The common commands are taken from the IEEE 488.2 (IEC 625-2) standard. A par-
ticular command has the same effect on different devices. The headers of these com-
mands consist of an asterisk "*" followed by three letters.
●
on page 74
*CLS
●
on page 74
*ESE
●
on page 74
*ESR?
●
on page 75
*IDN?
●
on page 75
*OPC
●
on page 75
*OPT?
●
on page 75
*RST
●
on page 75
*STB?
●
on page 75
*WAI
*CLS
Clear status
Sets the status byte (STB), the standard event register (ESR) and the EVENt part of
the QUEStionable and the OPERation registers to zero. The command does not
alter the mask and transition parts of the registers. It clears the output buffer.
Usage:
Setting only
*ESE <Value>
Event status enable
Sets the event status enable register to the specified value. The query returns the con-
tents of the event status enable register in decimal form.
Parameters:
<Value>
Range:
*ESR?
Event status read
Returns the contents of the event status register in decimal form and subsequently
sets the register to zero.
Return values:
<Contents>
Range:
Usage:
Query only
Allice Messtechnik GmbH
Remote Control - Commands
0 to 255
0 to 255
Remote Control