Measurement Commands - R&S NGE-K101 User Manual

Table of Contents

Advertisement

www.allice.de
®
R&S
NGE100B

6.5.10 Measurement Commands

User Manual 5601.1343.02 ─ 02
[SOURce:]POWer:PROTection:CLEar
This command resets the OPP state of the selected channel. If the OPP has tripped
the red blinking OPP message on the display will be cleared for the selected channel.
Usage:
Event
MEASure[:SCALar]:CURRent[:DC]?
This command queries the measured current value of the previous selected channel.
Return values:
<Current>
<numeric_value>
Return the measured current.
Default unit: A
Example:
MEAS:CURR? -> 1.000
Usage:
Query only
MEASure[:SCALar]:POWer?
This command queries the measured power value of the previous selected channel.
Return values:
<Power>
<Numeric_value>
Return the measured power value.
Default unit: W
Example:
MEAS:POW? -> 3.00E+00
Return the measured power value of the selected channel.
Usage:
Query only
MEASure[:SCALar][:VOLTage][:DC]?
This command queries the measured voltage value of the previous selected channel.
Return values:
<Voltage>
<numeric_value>
Return the measured voltage.
Default unit: V
Example:
MEAS? -> 1.000E+00
Usage:
Query only
Allice Messtechnik GmbH
Remote Control
Remote Control - Commands
88

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nge102bNge103bNge100bNge-k102Nge-k103

Table of Contents