R&S NGE-K101 User Manual page 84

Table of Contents

Advertisement

www.allice.de
®
R&S
NGE100B
User Manual 5601.1343.02 ─ 02
Example:
OUTP:MODE? -> CV
Output mode has changed to constant voltage mdoe.
Usage:
Query only
OUTPut:SELect <arg0>
This command activates or queries the instrument channel state of the previous
selected channedl.
If the selected channel is activated, the respective channel key (i.e. Ch 1, Ch 2, Ch 3)
on the front panel will be illuminated.
Parameters:
<arg0>
{OFF | ON | 0 | 1}
ON | 1
Selected channel will be activated.
OFF | 0
Selected channel will be deactivated.
*RST:
Example:
INST OUT1
OUTP:SEL ON
Ch 1 is activated.
OUTP:SEL? -> 1
Return the state of the selected channel.
OTP:TRIPped?
Queries the overtemperature protection (OTP) tripped state of the previous selected
channel.
It is tripped if an OTP event occurs. Available from V1.50
Return values:
<arg0>
{OFF | ON | 0 | 1}
ON | 1
OTP event occurs.
OFF | 0
OTP event not occurs.
*RST:
Example:
OUTP:TRIP? -> 1
OTP event occurs.
Usage:
Query only