Easyramp Commands - R&S NGE-K101 User Manual

Table of Contents

Advertisement

www.allice.de
®
R&S
NGE100B

6.5.12 EasyRamp Commands

User Manual 5601.1343.02 ─ 02
Parameters:
<Voltage>
{<Voltage> | MINimum | MAXimum | DEFault}
<Voltage>
0.0 V to 32.2 V (adjustable in 10 mV steps)
MINimum
0.000E+00
MAXimum
32.200E+01
Default unit: V
Example:
VOLT 10
Set the voltage value to 10 V.
VOLT? -> 10
Return the voltage value of the selected channel.
[SOURce:]VOLTage:EASYramp[:STATe] <arg0>
This command activates/deactivates or queries the EasyRamp state for the previous
selected channel.
See also SOURce:VOLTage:RAMP:STATe.
Parameters:
<arg0>
[SOURce:]VOLTage:EASYramp:TIME <Time>
This command sets/queries the EasyRamp time of the previous selected channel in
milliseconds. 10 ms to 10 s in steps of 10 ms.
See also SOURce:VOLTage:RAMP:DURation.
Parameters:
<Time>
Range:
*RST:
Default unit: ms
[SOURce:]VOLTage:RAMP:DURation <Time>
This command sets/queries the EasyRamp duration of the previous selected channel
in mili seconds. 10ms to 10 seconds in steps of 10 ms.
Parameters:
<Time>
Range:
*RST:
Default unit: ms
Allice Messtechnik GmbH
Remote Control - Commands
10 to 10000
10
10 to 10000
10
Remote Control
90

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nge102bNge103bNge100bNge-k102Nge-k103

Table of Contents