R&S NGE-K101 User Manual page 49

Table of Contents

Advertisement

www.allice.de
®
R&S
NGE100B
User Manual 5601.1343.02 ─ 02
Trigger out parameters Trigger conditions
Output Off
Fuse Tripped
CC Mode
Voltage Level
Current Level
Critical Event
EasyArb Active
Logic
Active High
Active Low
1. Select "Interface > Digital IO" menu item.
The R&S NGE100B displays the "Digital IO" dialog. See
Figure 5-19: Digital IO dialog
2. Select the desired DIO interface.
Allice Messtechnik GmbH
Description
Output the selected logic level when the
Output is turned off at the selected channel.
Output the selected logic level when a fuse
tripped event occurs on the selected chan-
nel.
See
Chapter 5.4.3, "Activating
on page 37 and
on page 42.
Output the selected logic level when the
selected channel operates in the CC mode.
See
"CC mode"
>= "set value"
Output the selected logic level when the
voltage level of the selected channel is
greater or equal to the set level.
>= "set value"
Output the selected logic level when the
current level of the selected channel is
greater or equal to the set level.
OVP
Output the selected logic level when the
selected critical event (OVP, OPP, OTP)
OPP
occurs on the selected channel. See
ter 5.5.1,
"Protection",
OTP
Output the selected logic level when the
EasyArb function is running on channel 1.
See
Chapter 5.5.3,
Note: For a reliable DIO trigger output, the
EasyArb waveform needs to be at least 200
ms.
Set the logic level of the trigger out signal.
Figure
Instrument Functions
Advanced Functions
Fuse",
Chapter 5.5.2,
"Fuse",
on page 35.
Chap-
on page 40.
"EasyArb",
on page 43.
5-19.
47