Example:
INST OUT1
OUTP ON
Ch 1 and output state will be activated; i.e. Ch 1 and Output key
on the front panel will be illuminated.
OUTP? -> 1
Return the output state.
OUTPut:GENeral <arg0>
This command activates or queries the instrument master output state.
If the master output state is activated, the Output key on the front panel will be illumi-
nated.
Parameters:
<arg0>
{OFF | ON | 0 | 1}
ON | 1
Output state will be activated
OFF | 0
Output state will be deactivated
Example:
INST OUT1
OUTP:SEL ON
Volt 12
Curr 0.1
Ch 1 is activated.
INST OUT2
OUTP:SEL ON
Volt 12
Curr 0.2
Ch 2 is activated.
OUTP:GEN ON
Ch 1 and Ch 2 and Output state are activated.
OUTP:GEN? -> 1
Return output state of the channel.
OUTPut:MODE?
Queries the output operating mode of the previous selected channel
Return values:
<arg0>
{OFF | CV | CC}
OFF
Output is turned off.
CV
Output mode changed to constant voltage mode.
CC
Output mode changed to constant current mode.
*RST:
Allice Messtechnik GmbH
Remote Control - Commands
OFF
Remote Control
81