FUSE[:STATe] <arg0>
This command defines or queries the fuse for the previous selected channel.
Parameters:
<arg0>
Example:

6.5.8 OVP Setting

[SOURce:]VOLTage:PROTection:MODE <arg0>
This command defines or queries the OVP mode for the previous selected channel.
Parameters:
<arg0>
Example:
User Manual 5601.1343.02 ─ 04
3
channel Ch 3
INST OUT1
FUSE:LINK 2
FUSE:UNL 2
Fuse Ch 1 is unlinked with Ch 2
Setting only
{0 | 1}
1
Fuse will be activated.
0
Fuse will be deactivated.
*RST:
OFF | 0
INST OUT1
FUSE ON
Fuse of Ch 1 is activated
FUSE? -> 1
Return fuse state of the selected channel.
MEASured | PROTected
MEASured
The OVP switches off if the measured value exceeds the thresh-
old.
PROTected
If the adjusted threshold is exceeded the output of the instru-
ment will be not switched on; additionally the measured value is
monitored (please also refer to function MEASured).
INST OUT1
VOLT:PROT:MODE PROT
Sets OVP protected mode for Ch 1.
VOLT:PROT:MODE? -> PROT
Return the selected channel OVP mode.