R&S
NGE100B
7.3.4 Combined voltage and current settings
The APPLy subsystem provides a command that enables you to set the current and
voltage of a channel in one step.
APPLy <arg0>
This command defines or queries the voltage and current value of the selected chan-
nel.
Parameters:
<arg0>
Example:
7.3.5 Output setting
The OUTPut subsystem contains the commands for activating the output channels.
User Manual 5601.1343.02 ─ 09
{<Voltage> | DEF | MIN | MAX} [, {<Current> | DEF | MIN | MAX}]
<Voltage>
Set voltage in the range of 0 V to 32.20 V.
DEF
Set voltage to default.
MIN
Set voltage to minimum.
MAX
Set voltage to maximum.
<Current>
Set current in the range of 3 mA to 3.025 A.
DEF
Set current to default.
MIN
Set current to minimum.
MAX
Set current to maximum.
INST OUT1
APPLY "6,2"
Set channel 1 to output 6 V and 2 A.
APPLY? -> 6, 2
Return voltage and current value of the selected channel.
Remote control commands
Configuration commands
80