Example:

7.3.3 Current setting

The SOURce:CURRent subsystem contains the commands for setting the current limit
of the output channel. The default unit is A.
Example: Configuring the current output
// ************************************************
// Select the channel
// ************************************************
INST OUT1
// ************************************************
// Set the current value
// ************************************************
// selects a channel and sets the current
CURR 2
// queries the current of the selected channel
CURR?
// response: 2.000
// ************************************************
// Query the range of the current values
// ************************************************
// queries the upper and lower limit of the current
CURR? MIN
// response: 0.003
CURR? MAX
// response: 3.025
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <Current>
This command defines or queries the current value of the selected channel.
Parameters:
<Current>
Example:
User Manual 5601.1343.02 ─ 09
MAXimum
32.200E+01
Default unit: V
VOLT 10
Set the voltage value to 10 V.
VOLT? -> 10
Return the voltage value of the selected channel.
<Current>
Set the selected channel to a current range of 3 mA to 3.025 A.
Default unit: A
CURR 2
Set the selected channel current to 2 A.
CURR? -> 2
Return current value of the selected channel.
Remote control commands
Configuration commands
79