R&S NGE100B User Manual page 86

Hide thumbs Also See for NGE100B:
Table of Contents

Advertisement

®
R&S
NGE100B
Example:
FUSE:TRIPped?
This command queries the fuse tripped of the previous selected channel.
Return values:
<arg0>
Example:
Usage:
FUSE:UNLink <arg0>
This command unlinks the channel fuses (fuse linking) for the previous selected chan-
nel.
Parameters:
<arg0>
Example:
Usage:
FUSE[:STATe] <arg0>
This command defines or queries the fuse for the previous selected channel.
Parameters:
<arg0>
User Manual 5601.1343.02 ─ 09
INST OUT1
FUSE:LINK 2
Fuse Ch 1 is linked with Ch 2.
FUSE:LINK? 2 -> 1
If the fuse of Ch 1 is linked with Ch 2, a "1" is returned.
FUSE:LINK? 3 -> 0
If the fuse of Ch 1 is not linked with Ch 3, a "0" is returned.
ON
Fuse is tripped.
OFF
Fuse is not tripped.
INST OUT1
FUSE:TRIP? -> 0
Fuse of Ch 1 has not tripped.
Query only
1
Channel Ch 1.
2
Channel Ch 2.
3
Channel Ch 3.
INST OUT1
FUSE:LINK 2
FUSE:UNL 2
Fuse Ch 1 is unlinked with Ch 2.
Setting only
{0 | 1}