Voltage Setting - R&S NGE100B User Manual

Hide thumbs Also See for NGE100B:
Table of Contents

Advertisement

®
R&S
NGE100B
Example:

7.3.2 Voltage setting

The SOURce:VOLTage subsystem contains the commands for setting the voltage of
the output channel. The default unit is V.
Example: Configuring the output voltage
This example contains all commands to configure and query the output voltage.
// ************************************************
// Select the channel
// ************************************************
INST OUT1
// ************************************************
// Set the voltage value
// ************************************************
// selects a channel and sets the voltage
VOLT 10
// queries the output voltage of a channel
VOLT?
// response: 10.000
// sets the voltage to maximum or minimum respectively
VOLT MAX
VOLT MIN
// ************************************************
// Query the range of the voltage values
// ************************************************
// queries the upper and lower limit of the output voltage
VOLT? MIN
// response: 0.000
VOLT? MAX
// response: 32.20
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <Voltage>
This command defines or queries the voltage value of the selected channel.
Parameters:
<Voltage>
User Manual 5601.1343.02 ─ 09
INSTrument OUTPut1
INSTrument OUTP1
Selects the output channel as Ch 1.
INST? -> 2
Return the output channel as Ch 2.
{<Voltage> | MINimum | MAXimum | DEFault}
<Voltage>
0.0 V to 32.2 V (adjustable in 10 mV steps)
MINimum
0.000E+00
Remote control commands
Configuration commands
78