Otherwise, the address must be set manually. Automatic Private IP Addressing
(APIPA) is not supported.
By default, the instrument is configured to use dynamic TCP/IP configuration and
obtain all address information automatically. This means that it is safe to establish a
physical connection to the LAN without any previous instrument configuration.
Risk of network errors
Connection errors can affect the entire network. If your network does not support
DHCP, or if you choose to disable dynamic TCP/IP configuration, you must assign valid
address information before connecting the instrument to the LAN. Contact your net-
work administrator to obtain a valid IP address.
Access:
1. Connect a LAN cable to the
2. Configure the connection settings manually or have the network configure the set-
tings for you via menu. By default, automatic configuration is set in the instrument.
For manual configuration, select "Ethernet" menu item
The R&S NGE100B displays the "Ethernet" dialog.
Figure 6-15: Ethernet dialog
3. Select "DHCP & Auto-IP" for "IP Mode".
4. Enter the "IP Address".
5. Enter the "Subnet Mask".
6. Enter the "Default Gateway".
7. Select "Apply" to complete the configuration.
8. Press [Menu] key to exit the dialog.