R&S NGE100B User Manual page 85

Hide thumbs Also See for NGE100B:
Table of Contents

Advertisement

®
R&S
NGE100B
FUSE:LINK 2
// queries the combined overcurrent protection of the selected channel
FUSE:LINK? 2
// response 1
// ************************************************
// Unlink linked overcurrent protection
// ************************************************
FUSE:UNLink 2
// queries the combined overcurrent protection of the selected channel
FUSE:LINK? 2
// response 0
FUSE:DELay <Delay>
This command defines and queries the fuse delay for the previous selected channel.
Parameters:
<Delay>
Example:
FUSE:LINK <arg0>
This command defines or queries the channel fuses (fuse linking) for the previous
selected channel.
Parameters:
<arg0>
User Manual 5601.1343.02 ─ 09
{<Delay> | MIN | MAX}
<delay>
0 s to 10 s (adjustable in 0.010 s steps e.g. FUSE:DEL 0.010 =
0.01 s)
MIN
Set fuse delay to minimum.
MAX
Set fuse delay to maximum.
Default unit: s
INST OUT1
FUSE:DEL 0.050
Set fuse delay of channel 1 to 50 ms.
FUSE:DEL? -> 0.05
Return the fuse delay time of the selected channel.
1
Channel Ch 1.
2
Channel Ch 2.
3
Channel Ch 3.