R&S
NGE100B
System test
"Beeper Test"
" Fan Test"
6.10.6 Updating the firmware
The latest firmware file can be found from the product homepage
schwarz.com/product/nge100b.
For firmware update using SCPI command, see
for version 1.3 and above"
Do not power off or operate the instrument during a firmware update.
Do not rename the firmware filename NGE100FWRelease.tar.gz.enc and
NGE100FWRelease.fwp.
1. Download the firmware file from the product homepage.
2. Unzip the firmware to the root directory of a USB flash drive.
There are two files after unzipping the firmware file, do not rename these files.
3. Connect the USB flash drive to the USB port of the instrument and wait for ten sec-
onds.
4. Press [Menu] key.
5. Select "Utilities" > "Update" menu item.
If firmware file is found, the R&S NGE100B displays the "Update" dialog. See
ure
Figure 6-27: Update dialog
6. Press [Enter] key to start firmware update process.
The R&S NGE100B reboots automatically as part of the update process.
7. Press [Enter] key to confirm the update when prompt.
User Manual 5601.1343.02 ─ 09
on page 69.
6-27.
Instrument functions
General instrument settings
Description
Test that the instrument sound is working. Three
beep sounds should be heard during the test.
Test that the cooling fan is working.
http://www.rohde-
"Firmware update via SCPI command
Fig-
68