Table 6-2: Trigger out parameters and conditions
Trigger out parameters Trigger conditions
Channel
Condition Trigger
Logic
After a recall function, the "Master Enable" is disabled. See
on page 53.
Enable the required DIO and "Master Enable" parameters to set up the Digital IO for
usage.
Access:
1. Press [Menu] key.
2. Select "Interface" > "Digital IO" menu item.
User Manual 5601.1343.02 ─ 09
Ch 1, Ch 2, Ch 3
Output On
Output Off
Fuse Tripped
CC Mode
Voltage Level
>= "set value"
Current Level
>= "set value"
Critical Event
OVP
OPP
OTP
EasyArb Active
Active High
Active Low
Instrument functions
Digital trigger I/O
Description
Output channel selected to monitor for trig-
ger condition.
Output the selected logic level when the
[Output] is turned on at the selected chan-
nel.
Output the selected logic level when the
[Output] is turned off at the selected chan-
nel.
Output the selected logic level when a fuse
tripped event occurs on the selected chan-
nel.
See
Chapter 6.3, "Electronic
on page 41.
Output the selected logic level when the
selected channel operates in the CC mode.
See
"CC mode"
on page 38.
Output the selected logic level when the
voltage level of the selected channel is
greater or equal to the set level.
Output the selected logic level when the
current level of the selected channel is
greater or equal to the set level.
Output the selected logic level when the
selected critical event (OVP, OPP, OTP)
occurs on the selected channel. See
ter 6.5,
"Protection",
on page 44.
Output the selected logic level when the
EasyArb function is running on channel 1.
See
Chapter 6.7.1,
"EasyArb",
Note: For a reliable DIO trigger output, the
EasyArb waveform needs to be at least 200
ms.
Set the logic level of the trigger out signal.
"Recall instrument settings"
fuse",
Chap-
on page 49.