274 | DR 800 | Problem solving

Error messages on the ceiling suspension

Table 51: Error messages are displayed on the display of the control panel of
the X-ray tube of the ceiling suspension
Error
message
FAULT EE-
PROM 567
FAULT EE-
PROM 628
SHORT
CIRC. 567
0392C EN 20210309 1049
Possible cause
Read Only Memory
(EEProm) of the PCB
567 of the ceiling
suspension unit main
CPU has stopped
working.
Read Only Memory
(EEProm) of the PCB
628 of the ceiling
suspension unit con-
trolbox has stopped
working.
Brake short circuit to
ceiling suspension.
The lateral SID is un-
known.
Possible solution
Stop use of the system and call technical
service. This may require a PCB replace-
ment.
Stop use of the system and call technical
service. This may require a PCB replace-
ment.
Stop use of the system and call technical
service. A hardware device has failed.
The ceiling suspension must be set into
one of three programmed positions to
display the lateral SID (with digital
switches).
If the SID does not display at one of the
positions, a mechanical switch may be
damaged, its connection wire discon-
nected, or it is out of alignment. The sys-
tem will only allow manual controls.
Contact technical service for support.