Radiographic Parameters for fluoroscopy

You can set up following radiographic parameters:
• kV: shows the radiographic kV value (X-ray tube voltage) for the exposure.
• mA: shows the radiographic mA value for the exposure.
When the automatic brightness system is on, an initial kV value is selected.
During the examination the radiographic parameters are adjusted
automatically and displayed in the software console.
Extra parameters for pulsed fluoroscopy
Following parameters are only available for pulsed fluoroscopy:
• ms: shows the pulse width (in milliseconds) selected for the exposure.
• FPS: frames per second
When the automatic brightness system is on, an initial kV value is selected.
During the examination the radiographic parameters are adjusted
automatically and displayed in the software console.