42 | DR 800 | Introduction

Emergency shutdown power switch

Use the emergency shutdown power switch, if a dangerous situation cannot be
eliminated by pressing the emergency stop button.
An installation must be fitted with two emergency stop buttons. They must be
effectively arranged so that the first button is close to the radiographic table
and easily reached by the operator. The second button must be in the operator
room and also easily reached by the operator. The emergency stop buttons are
installed and labeled by customer.
Figure 24: Example of a typical placement of the emergency stop buttons in
the examination room and in the operator room.
0392C EN 20210309 1049
WARNING:
Use the emergency shutdown power switch in case of danger to
patients, operators, third parties, or one of the units. The entire
system will be shut down and the power supply will be
disconnected. In this situation, the system must not be restarted
and the Agfa service organization must be promptly contacted.
WARNING:
It must be ensured that the emergency switches are always
freely accessible.
6