140 | DR 800 | Basic Workflow
12. Release the fluoroscopy pedal to stop the exposure.
The roadmapping sequence is stored and displayed as a roadmapping
sequence thumbnail in the lower half of the Image Overview pane. The
last image of the sequence is visible in the thumbnail.
A roadmapping sequence thumbnail is indicated with a transparant Play
icon in the center.
Figure 122: Thumbnail of the roadmapping sequence
13. Make as many roadmapping exposures as desired, using the same
roadmapping mask, by pressing the fluoroscopy pedal.
14. Click the roadmapping button in the software console to finish the
roadmapping workflow.
Figure 123: Roadmapping workflow is ended
The current roadmapping mask cannot be used any more to perform
another roadmapping examination.
15. To perform another roadmapping examination, using a new mask, start
another roadmapping workflow in the software console.
If multiple roadmapping workflows are performed, a filled or hollow
triangle at the bottom of the thumbnails provides a visual link between
roadmapping sequences that have been acquired using the same mask.
16. Perform quality control.
Roadmapping sequences can be post-processed to apply landmarking and
to adjust brightness and contrast.
17. If all images in the examination are OK, click Close and Send All.
If configured, the static images, roadmapping sequences and DSA
sequences are sent to the printer and/or PACS archive. The exam is placed
in the Closed Exams pane.
Fluo sequences are not sent to a PACS archive. To archive a selected
acquisition sequence, click the Store Sequence button before clicking
Close and Send All.
0392C EN 20210309 1049