74 | DR 800 | Introduction
to the patient), the significant zone of occupancy applies for operator and
staff.
2 2
Vertical axis indicating the operator position (at 60 cm distance from
1.
central X-ray beam) and height range of stray radiation measurement (0
cm - 200 cm)
Horizontal axis indicating the table height
2.
Stray radiation in mGy measured at the operator position
3.
Figure 53: Significant zone of occupancy and stray radiation during
examination with horizontal table
Table 3: Conditions for measuring stray radiation values represented in the
illustrations
Workflow
Object
SID
Exposure parameters 100 kV, 50 mA, 10 frames per second, 10 ms pulse
Total exposure time
Inherent filtration
0392C EN 20210309 1049
Warning: The radiation protection has to be applied to the
patient and to the operator.
1
1
Fluoroscopy
Anthropomorphic abdomen phantom
115 cm
width
100 s
2.9 mm Al equivalent
200 cm
200 cm
0 cm
0 cm
3