Remarks For Hf-Emission And Immunity (Radiographic Wall Stand Bucky With Moving Grid) - AGFA DR 800 User Manual

Table of Contents

Advertisement

Remarks for HF-emission and immunity
(radiographic wall stand bucky with moving grid)
Guidance and manufacturer's declaration – Electromagnetic emissions
The system is intended for use in the electromagnetic environment specified
below. The customer or the user of the system should assure that it is used in
Transmission Measure-
ments
RF emissions in accord-
ance with CISPR 11
RF emissions in accord-
ance with CISPR 11
Harmonic emissions in
accordance with IEC
61000-3-2
Voltage fluctuations /
flicker emissions in ac-
cordance with
IEC 61000-3-3
Guidance and manufacturer's declaration – Electromagnetic immunity
The system is intended for use in the electromagnetic environment specified
below. The customer or the user of the system should assure that it is used in
Immunity Test
Discharge of static
electricity in ac-
cordance with IEC
61000-4-2
Annex A
such an environment.
Agreement
Group 1
Class A
Not applicable
Not applicable
Annex B
such an environment.
IEC 60601-1-2
Test level
6 kV contact
8 kV air
DR 800 | Product Information | 319
Electromagnetic Environment
Guidelines
The system uses RF energy only
for its internal function. There-
fore, its RF emissions are very
low and are not likely to cause
any interference in nearby elec-
tronic equipment.
The system is suitable for use in
all establishments other than do-
mestic and those directly con-
nected to the public low-voltage
power supply network that sup-
plies buildings used for domestic
purposes.
Compliance
Electromagnetic
level
IEC 60601-1-2
Floors should con-
Test level
sist of wood, con-
crete or ceramic
tiles. The relative
humidity must be