b) Position the detector holder at the desired height.
c) Secure the position of the arm by turning the knob thoroughly
clockwise.
4. Mount the detector (or cassette) in the detector holder.
a) Loosen the two knobs to release the two adjustable clamps.
b) Widen the clamps enough to fit the detector size.
c) Insert the detector and close the clamps, making sure it is properly
secured and supported.
d) Fasten the clamps by turning the two knobs thoroughly clockwise.
Related Links
Mounting the lower limbs support, ankle support or head contention device
page 250
WARNING:
The arm is not equipped with mechanical stops. If it no
held manually, it may cause the detector holder to fall
onto the tabletop.
WARNING:
Hold the clamps on the outside to prevent crushing your
hands.