Guidance and manufacturer's declaration – Electromagnetic emissions
The system is intended for use in the electromagnetic environment specified
below. The customer or the user of the system should assure that it is used in
Transmission Measure-
RF emissions in accord-
ance with CISPR 11
RF emissions in accord-
ance with CISPR 11
Harmonic emissions in
accordance with IEC
61000-3-2
Voltage fluctuations /
flicker emissions in ac-
cordance with
IEC 61000-3-3
Guidance and manufacturer's declaration – Electromagnetic immunity
The system is intended for use in the electromagnetic environment specified
below. The customer or the user of the system should assure that it is used in
Immunity Test
Discharge of stat-
ic electricity in ac-
cordance with IEC
61000-4-2
0392C EN 20210309 1049
such an environment.
ments
Group 1
Class A
Not applicable
Not applicable
such an environment.
IEC 60601-1-2
Test level
8 kV contact
2/4/8/15 kV air
Annex A
Agreement
Electromagnetic Environment
The system uses RF energy only
for its internal function. There-
fore, its RF emissions are very
low and are not likely to cause
any interference in nearby elec-
tronic equipment.
The system is suitable for use in
all establishments other than
domestic those directly connec-
ted to the public low-voltage
power supply network that sup-
plies buildings used for domestic
purposes.
Annex B
Compliance
level
IEC 60601-1-2
Test level
Guidelines
Electromagnetic
Environment
Floors should consist
of wood, concrete or
ceramic tiles. The
relative humidity
must be at least

Hide quick links: