Starting the System

To start the system:
1. Switch on the electrical room switch.
Check that the emergency shutdown power switch for the system and the
emergency stop button for the radiographic table is not activated.
Check that the power switch of the DR 800 sync cabinet is on.
2. Start the NX workstation.
For detailed information about starting up NX, refer to the NX User
Manual, document 4420.
The NX application and the software console are available on the NX
workstation.
3. In a configuration with a wireless DR Detector, power on the DR Detector:
a) attach a fully charged battery pack to the DR Detector.
b) turn on the DR Detector.
c) if needed, register the DR Detector to the NX workstation.
For detailed information about starting up the DR Detector, refer to the DR
Detector User Manual.
Topics:
Warming-up of X-ray tube
Automated workflow for warming-up of X-ray tube
0392C EN 20210309 1049
WARNING:
No patient should be near or on the table until the system has
started and is in its starting position.