Radiated electro-
magnetic field
IEC 61000-4-3
Fast transient
electrical disturb-
ance variables /
bursts in accord-
ance with IEC
61000-4-4
Impulse voltages
(surges) in ac-
cordance with IEC
61000-4-5
Conducted RF in
acordance with
IEC 61000-4-6
3 V/m
80 MHz to 2.7
GHz
2 kV for power
supply lines
1 kV for input/
ouput lines > 3
m
0.5/1 kV differ-
ential mode
0.5/1/2 kV com-
mon mode
3 V
150 kHz to 80
MHz
6 V
ISM frequencies
DR 800 | Product Information | 313
30%, if the floor is
made of synthetic
material.
IEC 60601-1-2
Portable and mobile
Test level
RF communications
equipment should
be used no closer to
any part of the
equipment, includ-
ing cables, than the
recommended sepa-
ration distance cal-
culated from the
equation applicable
to the frequency of
the transmitter rec-
ommended separa-
tion distance.
IEC 60601-1-2
The quality of the
Test level
voltage supplied
should correspond
to a typical commer-
cial or clinical envi-
ronment.
IEC 60601-1-2
The quality of the
Test level
voltage supplied
should correspond
to that of a typical
commercial or clini-
cal environment.
IEC 60601-1-2
Portable and mobile
Test level
RF communications
equipment should
be used no closer to
any part of the
equipment, includ-
ing cables, than the
recommended sepa-
ration distance cal-
culated from the
equation applicable
to the frequency of
the transmitter rec-
ommended separa-
tion distance.
0392C EN 20210309 1049