Figure 162: Brightness curve
Region of interest
To optimize the image quality of the anatomatically relevant region in the
image, the automatic brightness system can be applied to a smaller region in
the center of the image. Only the brightness of the region of interest is taken
into account to control the exposure parameters.
To select another region of interest from the list of configured sizes, use the +
and - buttons. The size of the region of interest is displayed in cm.
Figure 163: Region of interest
Dose
Each of these buttons allows adjustment of the ABS dose (depending on
configuration at installation time). Each time a button is selected
(highlighted), the others are automatically deselected.
Table 38: Automatic filter
Dose
Fluoroscopy boost
To temporarily increase the dose while performing a fluoroscopy sequence,
push the BOOST button. To return to the normal level, push the button again.
low dose
high dose
DR 800 | Software Console | 225
0392C EN 20210309 1049