90 | DR 600 | Safety Directions
Safety Directions for the Ceiling Suspension
If the radiographic table is installed in the movement range of the ceiling
suspension, make sure that the X-ray tube, collimator or X-ray tube arm do not
collide with the tabletop, especially when the X-ray tube moves under the
tabletop.
If the radiographic wall stand is installed in the movement range of the ceiling
suspension, make sure that the X-ray tube, collimator or X-ray tube arm do not
collide with the radiographic wall stand.
3261K EN 20210708 0808
DANGER:
In case of uncontrolled movement of the ceiling suspension,
press the nearest emergency stop button and contact your local
service organization.
DANGER:
Make sure that no persons or objects are within the movement
area of the system where they can collide with moving parts of
the system.
WARNING:
Monitor the patient position (hands, feet, fingers, etc.) with
special care to avoid injury to the patient caused by unit
movements. Patient hands must be kept away from mobile
components of the unit. Intravenous tubing, catheters and other
patient connected lines should be routed away from moving
equipment.
WARNING:
Make sure that no clothes of the patient or the operator get
caught by moving parts of the system.