Compressor Cannot Be Released - AGFA DR 800 User Manual

Table of Contents

Advertisement

264 | DR 800 | Problem solving

Compressor cannot be released

De-
tails
Cause
Brief
Solu-
tion
0392C EN 20210309 1049
A patient is lying on the table and the compressor is lowered, but
cannot be released using the position controls.
Power failure.
System failure that prevents the remote control of the compressor.
1. Fetch the wrench.
The wrench is attached to the cable support turret.
Figure 176: Cable support turret with wrench attached to it
2. Turn the bolt on the rear side of the compressor carriage coun-
terclockwise (seen from the rear).
The compressor moves upward as the bolt is turned.
Figure 177: Rear view of the compressor carriage