102 | DR 800 | Safety Directions

Preventing damage to the equipment

Before starting any movement, make sure there are no obstacles in the
movement area.
No object or material must be placed on the tabletop, the image receptor or
the bars. While the image receptor or tabletop are moving, these objects could
be crushed.
Do not place any objects on any other surface. During tilting, such objects
could fall, causing damage or injury.
Do not climb up on any part of the tabletop cover or the image receptor. The
covers could be damaged and create problems or malfunctions.
Do not sit on the column or the X-ray tube arm when the equipment is in the
vertical position. The load could break the material and damage it.
Do not place glasses for contrast liquids or other containers on the equipment,
the console or the cabinet. The contrast liquid could penetrate some
components of the equipment causing malfunctions or incorrect
interpretation of the X-ray images. We suggest that you immediately eliminate
contrast liquid stains.
The position controls prevent unintended movement and stop the movement
as soon as the operator releases the controls.
The low ceiling sensor guarantees that the system will never collide with the
ceiling of the examination room in which it is installed. The sensor is optional
and gets enabled in case that the software for low ceiling limits fails to work. If
the sensor is triggered, a service intervention is needed to reactiate the
system.