AGFA DR 800 User Manual page 317

Table of Contents

Advertisement

Tests of Re-
IEC
sistance to
60601-1-2:2014
Disruption
Test Level
Conducted
3 V
RF
150 kHz to 80 MHz
IEC
61000-4-6
Radiated RF
3 V/m
IEC
80 MHz to 2.5 GHz
61000-4-3
DR 800 | Product Information | 317
Level of
Electromagnetic Environ-
Agree-
ment
ment
3 V
Portable and mobile RF com-
munications equipment
should be used no closer to
any part of the device, includ-
ing cables, than the recom-
mended separation distance
3 V/m
calculated from the equation
Spot fre-
applicable to the frequency of
quencies
the transmitter.
Recommended separation
distance:
d = 1.2
MHz
d = 1.2
MHz
d = 2.3
GHz
where 'P' is the maximum out-
put power rating of the trans-
mitter in watts (W) according
to the transmitter manufac-
turer and 'd' is the recommen-
ded separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, should be less than
the compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
150 kHz to 80
80 MHz to 800
800 MHz to 2.5
0392C EN 20210309 1049