Error Messages - AGFA DR 100e User Manual

Hide thumbs Also See for DR 100e:
Table of Contents

Advertisement

Error messages

Details on the contents of messages can be found in the service
documentation which is available to service personnel.
The device must be restarted in order to continue working.
Message
FAULTY POWER
ERROR V3
FILAMENT
ERROR V2
LOCKED STARTER
FAULTY CHOPPER
X-RAY ERROR
MAX. TIME
DATA ERROR
DR 100e (Film/CR) | Introduction | 27
Description
Error on the Charger or
Chopper. No energy
available.
Absent V3 power sup-
ply.
No filament current.
Absent V2 power sup-
ply.
Error during the start-
ing up.
Error in the chopper
The kV value has not
reached 75% of the re-
quested value within
the first 10ms of expo-
sure.
The maximum exposure
time has been reached.
Memory error, data
checksum error.
Action
Switch off, wait some
minutes, switch on, if
the error appears gain
call Service.
Switch off, wait some
minutes, switch on, if
the error appears gain
call Service.
Switch off, wait some
minutes, switch on, if
the error appears gain
call Service.
Switch off, wait some
minutes, switch on, if
the error appears gain
call Service.
Press the reset button
and repeat the expo-
sure.
Press the reset button
and repeat the expo-
sure.
Press the reset button
and repeat the expo-
sure.
Press the reset button
and repeat the expo-
sure.
Switch off, wait some
minutes, switch on, if
the error appears gain
call Service.
0360B EN 20181015 0853

Advertisement

Table of Contents
loading

This manual is also suitable for:

6012/0016012/002

Table of Contents