AGFA DR 800 User Manual page 320

Table of Contents

Advertisement

320 | DR 800 | Product Information
Fast transient elec-
trical disturbance
variables / bursts
in accordance with
IEC 61000-4-4
Impulse voltages
(surges) in accord-
ance with IEC
61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
iput lines in ac-
cordance with IEC
61000-4-11
Magnetic field at
the supply fre-
quency (50/60
Hz) in accordance
with IEC
61000-4-8
• REMARK : U
cation of the test level.
0392C EN 20210309 1049
2 kV for power
supply lines
1 kV for input/
ouput lines > 3 m
1 kV differential
mode
2 kV common
mode
< 5% U
for 0.5
T
cycle
(> 95% voltage
dip in U
T
40% U
for 5 cy-
T
cles
(60% voltage dip
in U
T
70% U
for 25 cy-
T
cles
(30% voltage dip
in U
T
< 5% U
for 5 s
T
3 A/m
is the alternating current in the network before the appli-
T
at least 30%, if the
floor is made of
synthetic materi-
al.
IEC 60601-1-2
The quality of the
Test level
voltage supplied
should corre-
spond to a typical
commercial or
clinical environ-
ment.
IEC 60601-1-2
The quality of the
Test level
voltage supplied
should corre-
spond to that of a
typical commer-
cial or clinical en-
vironment.
IEC 60601-1-2
Mains power qual-
Test level
ity should be that
of a typical com-
mercial or hospi-
tal environment.
If the user of the
system requires
continued opera-
tion during power
mains interrup-
tions, it is recom-
mended that the
system be pow-
ered from an un-
interruptible pow-
er supply or a bat-
tery.
IEC 60601-1-2
Magnetic field at
Test level
the network fre-
quency should
correspond to the
typical values as
they are in a com-
mercial and clini-
cal environment.