Performing offset calibration

Figure 151: Offset Calibration button
The Offset Calibration button reminds the user to manually trigger the offset
calibration in case the system cannot perform the offset calibration
automatically.
It is required to manually trigger the offset calibration, in order to prevent
image artifacts caused by delayed offset calibration.
1. Press the offset calibration button in the DR Detector switch.
The X-ray modality controls are displayed on the software console.
2. Press the offset calibration button.
The offset calibration is performed.
DR 800 | Software Console | 195
0392C EN 20210309 1049