Prepare The Full Leg Full Spine Configuration - AGFA DR 800 User Manual

Table of Contents

Advertisement

Prepare the Full Leg Full Spine configuration

Position the X-Ray system and the FLFS overlay:
1. Place the tabletop in the horizontal position.
2. Position the FLFS overlay on the radiographic table.
a) Insert the mounting unit on the tabeltop guide.
If the patient is positioned with the head on the left side, the mounting
unit must be inserted on the rear guide. If the patient is positioned
with the head on the right side, the mounting unit must be inserted on
the front guide.
b) Fasten the mounting unit by turning both the handle and the smaller
knob thoroughly clockwise.
Figure 124: Fastening the mounting unit (head left)
c) Put the overlay on the tabletop and slide it into the mounting unit until
it automatically fixes.
Figure 125: Mounting the overlay (head left)
NX is configured for a specific patient orientation, either head left
(default) or head right.
In longitudinal direction the position of the overlay on the tabletop can be
chosen depending on the reqion of interest of the planned examination.
The overlay should not protrude from the sides of the tabletop.
2
4
DR 800 | Full Leg Full Spine | 161
1
3
3
2
4
0392C EN 20210309 1049
4

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6010/200

Table of Contents