DR workflow with fluoroscopy for positioning

This workflow is available only on DR systems that support dynamic imaging.
Fluoroscopy can be used as a guide for positioning the patient before
performing the planned exposure.
To use fluoroscopy for positioning:
1. Add a fluo group to the Image Overview pane.
If a fluo group has already been added based on data from the RIS, this
step can be skipped.
a) In the Examination window, click Add Image.
The Add Image window appears.
Figure 62: Add Image
b) Specify the exam group and exam type by clicking on the buttons.
c) Select an exam type that is configured as a fluo group and click OK.
The fluo group thumbnail is added to the Image Overview pane.
A fluo group thumbnail is indicated with an icon in the top right corner of
the thumbnail.
Figure 63: Thumbnail for a fluo group
2. Select the thumbnail for the fluo group in the Image Overview pane of
the Acquisition window.
The selected DR detector is activated. The default X-ray exposure
parameters and X-ray system position for the selected examination are
sent to the modality. The software console displays these settings in the
examination overview.