DR 800 | Software Console | 201

DAP Value

The DAP value shows the radiation value of the last exposure. The radiation
2
measure is read as DAP value (Dose Area Product) in cGy*cm
(for example:
DAP 12.22).
A new exposure resets the DAP value.
Dose measurement values
When the screen with generator controls is displayed, more dose
measurement values become visible:
2
• DAPR (Dose Area Product Rate) in cGy*cm
/s
• AK (Air Kerma) in mGy (calculated value)
• AKR (Air Kerma Rate) in mGy/min (calculated value)
Keep in mind that the Air Kerma value is not correct when using the optional
ceilling suspension with extra filtration.
0392C EN 20210309 1049