268 | DR 800 | Problem solving
Stitching fails
Table 49: Problem: Stitching fails
Details
Cause
Brief Solution
0392C EN 20210309 1049
The partial images cannot be stitched because the grid
markers in the stitching grid are not visible in the parti-
al images.
The stiching grid was not used for the examination.
Click Cancel in the Stitching pane. No DR Full Leg Full
Spine image is available.