106 | DR 800 | Safety Directions

Ceiling suspension and radiographic wall stand

0392C EN 20210309 1049
WARNING:
All movements of the ceiling suspension are blocked by
permanent magnet brakes that require power to be turned ON to
release them. This means that in the event of accidental
interruption of power, the brake remains active, so one must
avoid manually force the movement of the device.
WARNING:
Keep a firm grip on the handle of the X-Ray tube head unit
whenever activating the brakes, to keep the tube housing under
full control.
WARNING:
Use the brakes correctly. Activate the positioning controls to
move the equipment. Release the positioning controls only after
the equipment is set in place.
WARNING:
In the bucky and in the bucky drawer there are metal
protrusions and pinch points that can cause damage to the
user's hands. So use only the proper handle to pull out and
insert the cassette tray from the bucky.
WARNING:
Do not try to force the X-ray tube towards the bottom when the
main rope is broken! Never use the ceiling suspension after the
safety device intervention. Immediately contact service and
replace the broken rope.
WARNING:
If the wall stand unit is used with a mobile table or stretcher, it is
important that the operator must block the wheels to firmly
position the stretcher while helping the patient on or off the
table.
CAUTION:
Do not use the wallstand if it shows damage to one of the two
ropes or obvious signs of breakage or wear. Contact service to
replace broken ropes.