Collimation
cm
kV
43x43
100
43x43
100
43x43
100
43x43
100
43x43
100
35x43
105
35x43
105
43x43
110
43x43
110
43x43
110
* Radioscopy (continuous and pulsed) used in dynamic examinations (e.g.,
urography, cistography, clisma barium, digestive and alimentary tract)
** Values typical of skeleton examinations (e.g., skull, upper and lower
extremities, spine, pelvis)
*** Values typical of abdomen and thorax examination
The addition of a copper filter is typical of paediatric examination.
It is common knowledge that the ionization process can cause damages to
cellular DNA. See following information about deterministic effects on human
body after localised and whole body irradiation by X-ray (data source: ICRP,
International Commission on Radiological Protection):
Organ/
Effect
Tissue
Testicles
Temporary sterility
Permanent sterility
Ovaries
Sterility
Ocular lens Detectable opacities
Visual impairment
(cataract)
Bone mar-
Haemopoiesis impair-
row
ment
Skin
1. Erythema
2. Desquamation
RADIOGRAPHY FFD 180 cm (***)
0Al
Ak
Collimation
mAs
mGy
cm
1
0,048
25x30
2
0,092
30x30
6,3
0,270
30x30
10
0,420
43x43
20
0,977
43x43
1,6
0,148
43x43
10
0,681
43x43
2
0,109
43x43
6,3
0,329
43x43
10
0,509
43x43
THRESHOLD AB-
SORBED DOSE
(Gy)
Short-term expo-
sure
Single dose
0.15
3.5 - 6.0
2.5 - 6.0
0.5 - 2.0
5.0
0.5
2
18
DR 800 | Product Information | 305
1Al - 0,2 Cu
Ak
kV
mAs
mGy
70
2,5
0,015
90
2
0,023
90
10
0,132
100
1
0,020
100
2
0,038
100
6,3
0,113
100
10
0,177
110
2
0,049
110
6,3
0,151
110
10
0,233
Long-term
exposure
Yearly repea-
ted for many
years
0.4
2.0
> 0.2
> 0.1
> 0.15
> 0.4
-
-
0392C EN 20210309 1049