134 | DR 800 | Basic Workflow
Figure 110: Icon indicating that injection of the contrast medium can start
9. Start injecting the contrast medium when the syringe icon is shown.
Blood vessels that contain the contrast medium, become clearly visible.
10. Release the radiography pedal or the exposure button to stop the
exposure.
The DSA sequence is stored and displayed as a DSA sequence thumbnail in
the lower half of the Image Overview pane. The last image of the
sequence is visible in the thumbnail
A DSA sequence thumbnail is indicated with a transparant Play icon in the
center.
Figure 111: Thumbnail of a DSA sequence
11. Perform quality control.
DSA sequences can be post-processed to change the mask image, to apply
pixel shift to correct for movement in the subtracted image, or to apply
landmarking by making anatomical background as visible as desired.
By modifying the MUSICA settings of the image, the presentation of the
DSA sequences can be fine-tuned.
12. If all images in the examination are OK, click Close and Send All.
If configured, the static images and DSA sequences are sent to the printer
and/or PACS archive. The exam is placed in the Closed Exams pane.
Fluo sequences are not sent to a PACS archive. To archive a selected
acquisition sequence, click the Store Sequence button before clicking
Close and Send All.
0392C EN 20210309 1049