c) Heating block
temperature control
is not working
correctly
8.5
Vacuum workstation-related errors
a) No vacuum is
received
b) Vacuum lost during
sample preparation
c) Filter probes not
working properly
d) Liquid left in some
wells in the
immobilization plate
e) White remains
(Sepharose beads)
in the
immobilization plate
8-6
Comments and suggestions
Remove and reinsert the heating block, see
Section 7.4.
If there is liquid between the heating block and
the instrument, clean the lens array, see Section
7.3.
If the problem persists, the contact pins may need
replacing. Contact QIAGEN Technical Services.
Comments and suggestions
Turn off the vacuum pump and open the cap to
the waste container to release any pressure. Close
the cap and start the pump again. Empty the
waste container if full.
Ensure that the tubing is connected correctly and
that there is no leakage.
Ensure that the tubing is connected correctly and
there is no leakage.
Ensure filter probes are working properly; see
Section 7.8.1.
Replace the corresponding filter probes; see
Section 7.8.2.
If more than 1 min has elapsed since the plate (or
strips) was agitated, agitate again for 1 min
before capturing the beads.