Maintenance
7.1

Shutdown procedure

Before carrying out maintenance, ensure the PyroMark Q96
ID is shutdown using the following procedure:
1. Choose "Instrument" from the main menu.
2. Choose "Shut Down Instrument".
3. Click "Yes" to confirm shutdown of the instrument. Error
4. All indicators on the front panel illuminate and the
5. Turn off the standby switch on the front of PyroMark Q96
7.2
Routine maintenance of the PyroMark
Q96 ID
The PyroMark Q96 ID should be checked and cleaned
regularly to remove dust and spillages.
Important points before starting:
The following items are required:
Procedure:
1. Shut down the instrument, see Section 7.1.
2. Open the instrument lid.
3. Clean the area around the dispensing unit and process
4. Open the process chamber lid. Clean the area around
7-2
messages will appear if the selected instrument is not
connected or the selected instrument is busy.
shutdown procedure begins. Shutdown is complete when
the message "Now it is safe to turn off instrument" is
displayed.
ID and check that the Power indicator lamp is turned off.
Avoid harsh cleaners and chemicals, and getting
moisture inside the instrument
The cleaning liquid must be applied to the cloth only
Ethanol (70%)
Clean, nonabrasive, lint-free cloths
chamber using a lint-free cloth moistened with 70%
ethanol.
the heating block.
PyroMark Q96 ID User Manual 01/2016