Moving the PyroMark Q96 ID

WARNING
Risk of personal injury and material damage
The PyroMark Q96 ID is too heavy to be lifted by one
person. To avoid personal injury or damage to the
instrument, do not lift the instrument alone.
If you need to transport the PyroMark Q96 ID, package the
instrument as follows:
1. When the instrument is not processing, shut down
2. Switch off the operator's computer and PyroMark Q96 ID
3. To move the instrument to a new location within the
4. Place the instrument onto a trolley and move to the new
5. Position the PyroMark Q96 ID on the bench. See Section
6. Follow the reinstallation procedures in Section 4.4 for
7. To move the instrument to another site: First, mount the
8. Pack the PyroMark Q96 ID and other components using
PyroMark Q96 ID User Manual 01/2016
PyroMark Q96 Software and exit from Microsoft
Windows.
and disconnect the mains power cable from the
instrument.
same laboratory, go to step 4. To move the instrument to
another site, go to step 7.
location.
Note: The instrument weighs 52 kg. Two persons are
required.
4.2 for site requirements.
stand-alone installation or Section 4.5 for network
installation.
transport protection block under the instrument lid.
the original packaging supplied with the instrument, if
available.
Note: The total weight of the package including the
instrument is 80 kg. The instrument alone weighs 52 kg.
Use suitable lifting equipment when moving the
package.
Installation Procedures
[W2]
4-3