Install a LAN system as follows:
1. Connect one end of the cross-connected Ethernet
2. Connect the other end of the network cable to the built-in
3. Connect the second Ethernet network cable between the
4. Connect mains power as per Section 4.6.
5. Start up the operator's computer and assign a unique IP
4.6

Connecting mains power

1. Ensure the standby switch on the front of the PyroMark
2. Connect the mains cable to the inlet on the rear of the
4.7

Setting up network properties

4.7.1

Used IP addresses for instrument and computer

The instrument IP address and the host IP address (operator's
computer) are fixed and do not need to be changed if the
PyroMark Q96 ID is connected to a LAN.
PyroMark Q96 ID User Manual 01/2016
network cable provided to the "NET" connector on the
rear panel of the PyroMark Q96 ID.
LAN outlet on the operator's computer.
Ethernet USB adaptor of operator's computer and the
nearest LAN connection point.
address to the second network adapter for communication
with the LAN.
Q96 ID is in position
switch).
PyroMark Q96 ID. Connect the mains cable to a
grounded mains power outlet. The PyroMark Q96 ID will
automatically adjust to the correct voltage.
Operator's computer IP address: 192.168.255.200
PyroMark Q96 ID IP address: 192.168.255.201
Installation Procedures