Procedure:
1. Ensure that no vacuum is applied to the Vacuum Prep
2. Disconnect the vacuum pump from the mains power.
3. Remove the broken tubing at one end and empty any
4. Disconnect the other end of the tubing and discard the
5. Cut the new vacuum tubing into three pieces and
6. Reconnect the vacuum pump to the mains power.
PyroMark Q96 ID User Manual 01/2016
Tool, i.e. the vacuum switch is closed (Off), and the
vacuum pump is switched off.
liquid waste into an empty beaker.
tubing and any liquid waste.
assemble it. Use the old tubing as a template for the size
of the new tubing pieces. Ensure that the tubing is
connected to the pump's vacuum fitting.
Maintenance
7-11