Installation Procedures
The PyroMark Q96 ID must be placed within approximately
1.5 m (59 in.) of a properly grounded (earthed) AC power
outlet. The power line to the instrument should be voltage
regulated and surge protected.
Note: It is recommend to plug the instrument directly into its
own wall socket and not to share the socket with other lab
equipment.
CAUTION
Risk of overheating
To ensure proper ventilation, maintain a minimum
clearance of 10 cm at the sides and rear of the PyroMark
Q96 ID.
Slits and openings that ensure the ventilation of the
PyroMark Q96 ID must not be covered.
Power requirements
The PyroMark Q96 ID operates at:
Depending on the unit purchased, the PyroMark Q96
Vacuum Workstation operates at:
Ensure that the voltage rating of the PyroMark Q96 ID and
PyroMark Q96 Vacuum Workstation is compatible with the
AC voltage available at the installation site. Mains supply
voltage fluctuations are not to exceed 10% of nominal supply
voltages.
Grounding requirements
To protect operating personnel, the PyroMark Q96 ID must
be correctly grounded (earthed). The instrument is equipped
with a 3-conductor AC power cord. To preserve this
protection feature, do not operate the instrument from an AC
power outlet that has no ground (earth) connection.
4-2
100–240 V AC, 50/60 Hz, 1.1–0.45 A.
100 V AC, 50/60 Hz, 1.7/1.4 A
115 V AC, 60 Hz, 1.5 A
230 V AC, 50 Hz, 0.6 A
PyroMark Q96 ID User Manual 01/2016